ProbeTurbidity digital sensor NTU for ODEON
RM0.00
◼ IR optical sensor with optical fibre
◼ Range : 0 to 4000 NTU or 0-4500 mg/L
◼ Robust and waterproof (IP68)
◼ Ultra low-power consumption
◼ Digital output Modbus RS-485
◼ Nephelometry measurement
Application :
◼ Urban wastewater treatment (inlet/ outlet controls)
◼ Sanitation network
◼ Industrial effluent treatment
◼ Surface water monitoring
◼ Drinking water
- Description
- Additional information
Description
- Technical Data
- Download
- Downloads
Optical technology :
The measure principle is based on IR nephelometry / 850 nm. The sensor can be calibrated with a formazine standard solution.
The NTU sensor integrates a low-cost optical technology, with a very few maintenance and no consumables.
Digital communication :
The PONSEL sensor can be connected to any types of transmitters, display units, controllers or data loggers with Modbus RS-485 or SDI-12 inputs. The optical sensor saves its calibration data for better measures management.
Integrated transmitter :
All data concerning calibration, history, users and measures are directly treated within the NTU sensor and transmitted via RS-485 or SDI-12.
Physical characteristics :
Compact, robust and light, the PVC sensor allows a handheld or fixed unit application.
Technical characteristics :
Measures | ||
Measure principle | Diffusion IR at 90° | |
Measure ranges |
0 to 4000 NTU in 5 ranges:
|
0 to 4500 mg/L Calibration : Range 0-500 mg/L according to NF EN 872 Range >500 mg/L according to NF T 90 105 2 |
Resolution | 0,01 to 1 NTU – mg/L | |
Accuracy | < 5% of the reading | |
Working temperature | 0°C to + 50°C | |
Measure of temperature | Via CTN | |
Stocking temperature | -10°C to + 60°C | |
Signal interface | Modbus RS-485 (standard) and SDI-12 (option) | |
Maximum refreshing time | < 1 second | |
Sensor power-supply | 5 to 12 volts | |
Electric consumption | Standby : 40 µA Average RS485 (1 measure/seconde) : 820 µA Average SDI12 (1 measure/seconde) : 4,2 mA Current pulse : 500 mA |
|
Sensor | ||
Dimensions | Diameter : 27 mm; length : 170 mm | |
Weight | 300 g (sensor + cable 3 meters) | |
Material | PVC, DELRIN, Quartz, PMMA, Polyamide | |
Maximum pressure | 5 bars | |
Connection | 9 armoured connectors, polyurethane jacket, bare-wires or waterproof Fisher connector | |
Degree of protection | IP68 |
Outline Drawing |
Wiring diagram |
||||
Cable length up to 15m |
Cable lenght 15 to 100 meters Red
V-
|
||||
1- Red | Power supply V+ | ||||
2 – Blue | SDI-12 | ||||
3 – Black |
Power supply V- | ||||
4 – Green |
B ” RS-485 “ | ||||
5 – White |
A ” RS-485 “ | ||||
6 – |
Cable shield |
Additional information
Cable Length | 3m, 7m, 15m |
---|