ProbeTurbidity digital sensor NTU for ODEON

SKU: N/A

RM0.00

◼ IR optical sensor with optical fibre
◼ Range : 0 to 4000 NTU or 0-4500 mg/L
◼ Robust and waterproof (IP68)
◼ Ultra low-power consumption
◼ Digital output Modbus RS-485
◼ Nephelometry measurement

Application :
◼ Urban wastewater treatment (inlet/ outlet controls)
◼ Sanitation network
◼ Industrial effluent treatment
◼ Surface water monitoring
◼ Drinking water

Clear
  • Description
  • Additional information

Description

  • Technical Data
  • Download
  • Downloads

Optical technology :
The measure principle is based on IR nephelometry / 850 nm. The sensor can be calibrated with a formazine standard solution.
The NTU sensor integrates a low-cost optical technology, with a very few maintenance and no consumables.
Digital communication :
The PONSEL sensor can be connected to any types of transmitters, display units, controllers or data loggers with Modbus RS-485 or SDI-12 inputs. The optical sensor saves its calibration data for better measures management.
Integrated transmitter :
All data concerning calibration, history, users and measures are directly treated within the NTU sensor and transmitted via RS-485 or SDI-12.
Physical characteristics :
Compact, robust and light, the PVC sensor allows a handheld or fixed unit application.

Technical characteristics :

Measures
Measure principle Diffusion IR at 90°
Measure ranges

0 to 4000 NTU in 5 ranges:

  • 0 – 50 NTU
  • 0 – 200 NTU
  • 0 – 1000 NTU
  • 0 – 4000 NTU
  • AUTOMATIC
0 to 4500 mg/L
Calibration :
Range 0-500 mg/L according to NF EN 872
Range >500 mg/L according to NF T 90 105 2
Resolution 0,01 to 1 NTU – mg/L
Accuracy < 5% of the reading
Working temperature 0°C to + 50°C
Measure of temperature Via CTN
Stocking temperature -10°C to + 60°C
Signal interface Modbus RS-485 (standard) and SDI-12 (option)
Maximum refreshing time < 1 second
Sensor power-supply 5 to 12 volts
Electric consumption Standby : 40 µA
Average RS485 (1 measure/seconde) : 820 µA
Average SDI12 (1 measure/seconde) : 4,2 mA Current pulse : 500 mA
Sensor
Dimensions Diameter : 27 mm; length : 170 mm
Weight 300 g (sensor + cable 3 meters)
Material PVC, DELRIN, Quartz, PMMA, Polyamide
Maximum pressure 5 bars
Connection 9 armoured connectors, polyurethane jacket, bare-wires or waterproof Fisher connector
Degree of protection IP68
Outline Drawing

Wiring diagram

 

Cable length up to 15m

 

Cable lenght 15 to 100 meters Red
Purple
Yellow         Power supply
Orange        V+ pink

  1. – Blue        SDI-12
  2. – Black       Power supply

V-

  1. – Green       B ” RS-485 “
  2. – White       A ” RS-485 “
  3. –            Cable shield Green/yellow
1- Red Power supply V+
2 – Blue SDI-12

3 – Black

Power supply V-

4 – Green

B ” RS-485 “

5 – White

A ” RS-485 “

6 –
Green/yellow

Cable shield

Additional information

Cable Length

3m, 7m, 15m