  • 4 electrodes (2 graphic, 2 platinum)
  • Range 0 to 200 mS/cm
  • Digital sensor / Modbus RS-485
  • Robust and Watertight
Applications :
  • Urban wastewater treatment
  • Industrial effluent treatment
  • Surface water monitoring
  • Sea water
  • Drinking water
Mounting at 4 electrodes:

The electrode works with a technology in 4 electrodes: an alternating current of constant-voltage is established between a primary’s pair of electrodes in graphite. The secondary’s electrodes in platinum allow of regulate the voltage imposed to primary’s electrodes to reflect of the fouling. The voltage measured between the primary’s electrodes is in function of the resistance of place and so, of the conductivity.

Digital Technology :

The “smart” Digital C4E sensor stores calibration and history data within the sensor. This allows you a “plug and play” system without re-calibration.

Thanks to the Universal Modbus RS485 protocol, the PONSEL Digital C4E can be connected to all devices commonly used (Datalogger, Controller, Automat, Remote System…).

Clear
  • Description
  • Additional information

Description

  • Technical Data
  • Learn more
  • Downloads
Measures
Measure principle Conductivity sensor with 4 electrodes (2 graphic, 2 platinum).
 

Measure ranges conductivity

0-200,0 µS/cm
0 –2000 µS/cm
0,00 –20,00 mS/cm
0,0 –200,0 mS/cm
Resolution Conductivity 0,01 to 1 according the range
Accuracy Conductivity +/- 1 % of the full range
Beyond 100 mS/cm use appropriate buffer solution
Measure range salinity 5-60 g/Kg
Measure range TDS -KCl 0-133 000 ppm
Measure range Temperature 0,00 °C to + 50,00°C
Resolution Temperature 0,01 °C
Accuracy temperature ± 0,5 °C
Response time < 5 s
Working temperature 0°C to 50°C
Temperature compensation NTC
Stocking temperature – 10°C to + 60°C
Signal interface Modbus RS-485 (option SDI-12)
Maximum refreshing time Max < 1 s
Sensor power-supply 5 to 12 volts
Electric consumption Standby : 25 µA
Average RS485 (1 measure/seconde) : 6,3 mA Average SDI12 (1 measure/seconde) : 9,2 mA
Current pulse : 500 mA
Sensor
Dimensions Diameter : 27 mm ; Lenght : 157 mm
Weight 350g (sensor + 3 m cable)
Material PVC, DELRIN, stainless steel
Maximum pressure 5 bars
Connection 9 armoured connectors, polyurethane jacket, bare-wires or waterproof Fisher connector
Protection IP68
Dimensions

Wiring diagram

 

 

Cable lenght 15 to 100 meters

Red Purple
Yellow        Power supply
Orange                V+ pink

  • – Blue        SDI-12
  • – Black       Power supply V-
  • – Green      B ” RS-485 “
  • – White      A ” RS-485 “
Cable length up to 15m 6 –           Cable shield Green/yellow

1- Red

Power supply V+

2 – Blue

SDI-12

3 – Black

Power supply V-

4 – Green

B ” RS-485 “

5 – White

A ” RS-485 “

6 –

Cable shield

Green/yellow

Never exceed a voltage of 10VDC (absolute maximum rating) on communication lines RS485, A or B, under penalty of irreversible destruction of the transceiver component RS 485.

SDI-12: respect the voltage value described in the associated standard (nominal: 5 VDC) Always connect ground + shield first.

None

Additional information

Cable Length

3m, 7m, 15m